Validate AI the way you validate a process.

Med Device is cautious about AI. We already have the method to control it.

Med Device is cautious about AI, and for good reasons. AI is seen as misunderstood and risky. Getting it implemented at all is the biggest hurdle. Privacy and data loss prevention are paramount to getting AI into use at any Med Device company. No data goes into a tool that has not been approved for it. Building a viable and usable product is the next hurdle, and that one is not exclusive to Med Device.

The caution makes sense. But AI works like any other human-driven process with multiple variables to control. We already know how to control a process like that. We validate it.

To be clear on scope. This is about AI tools used inside the quality system and the business. AI inside the device itself is a different path with its own regulatory requirements. Software validation requirements still apply to these tools, and the effort should follow the risk. IQ, OQ and PQ are how I think about the evidence.

Validation cannot be rigid

Validation of AI in Med Device cannot be rigid and deterministic. It has to match the way AI works. The same input can give a different output. That is variation, and we deal with variation every day.

Knowing AI is not deterministic, every output will have a pseudo tolerance, similar to a part coming out of production. Fuzzy matching will be needed in some cases. In others no deviation will be granted. Decide which is which before you start.

One passing run proves nothing. Run the same golden input multiple times and measure the pass rate. Set the sample size and the acceptance criteria before the run, the same way we do for any attribute test. For example, 59 passes with zero failures demonstrates 95% reliability with 95% confidence. The risk of the output sets the number.

It sometimes takes months to develop, validate and control a process. AI is no different. The better you build the process with AI, the more repeatable the output becomes.

IQ, OQ and PQ still apply

The stages we already use fit. So do the tools around them. The harness is the process being validated: the model plus the prompts, code and data around it.

StageFor an AI process
Plan Lay out the process, the requirements and the intended use. Define what a bad output looks like before anything is built.
DOE Test the code to understand the key parameters that affect the output.
Characterization Run the harness to understand the baseline inputs and outputs.
IQ Verify the harness is installed as specified. Model version pinned. Prompts and code under revision control. Data sources and access match the plan.
OQ Push the limits with difficult tests. Worst-case inputs, missing data and requests it should refuse.
PQ Run the harness at normal conditions, repeatedly, against acceptance criteria set in advance. PQ is essentially production hardening.
GOLDEN SET HARNESS EVAL PASS? RELEASE human written model + code the gage IQ · DEFINE PROCESS OQ · PUSH THE LIMITS PQ · PRODUCTION HARDENING FAIL → FIX → RE-RUN
Fig. 1 · The validation loop for an AI system

The eval is the gage

Capture the baseline by utilizing evals. A good eval is built on human-written golden data sets and it deterministically measures the output of the AI.

Start with deterministic evals. Where the output is free text and an exact match is not possible, a grader is needed. If that grader is AI, it is a gage, and it cannot be used until it passes an agreement study against human-graded examples. An unqualified AI grading AI is not a measurement.

The eval is our gage, and the eval is what gets qualified, not the AI. The engineer needs to understand how effective the eval is at measuring the harness. It has to recognize what good and bad look like, similar to understanding the alpha and beta risk of a gage. If the eval can incorrectly characterize an output, that is variation introduced into the process that isn't expected.

The eval is used to measure acceptance to launch a process into production. After launch it continues to control the process.

You cannot control what you cannot measure. same rule

What I have not proven yet

Production monitoring is the tricky part. During an eval you are running known golden inputs. During production the inputs will vary greatly. The deterministic eval needs to cover all known variation from the input.

A qualified AI grader may be part of the answer. The reviewer correction rate may be another. Every correction a human makes is a data point, and it can be trended on a control chart. I have not proven either out yet. Both are on my list.

Changes after launch

Post-launch, re-validation is triggered by any significant change that could affect the quality of the output. A model change. An update to the requirements in the process. Security updates. System architecture updates. Same as any other validated process.

Nothing ships without a human review

In Med Device, nothing can ship without a human review. The phrase "AI made a mistake" cannot occur. It is similar to how no product can be released without going through final verification and the product release process.

This looks different depending on the risk. For most internal AI tools, the output would go through multiple levels of cross-functional review and be approved with Part 11 compliant signatures.

In most scenarios AI should be viewed as a coworker. No quality record in Med Device is approved by its originator, and that principle applies to AI.

Human review is not a perfect control. We know inspection is not 100% effective, and a reviewer who sees good output all day will stop looking closely. So the review has to be designed like any inspection, with defined criteria, trained reviewers and a check on the reviewers.

The human still reviews. The time saved comes from reviewing a finished draft instead of starting from a blank page.

Why this is good news

If you are an engineer or a leader in a heavily regulated environment, you already have the method. Requirements, protocols, acceptance criteria, change control and review are the hard parts, and they are already in place. AI is the new part.

Just like anything else, you cannot control what you cannot measure. Evals are a critical step in controlling and optimizing AI processes.

This article describes a way of thinking. Software used in a quality system has its own validation requirements (for example ISO 13485 clause 4.1.6). Check your own procedures and regulators.

Compare notes.

Working on validating AI in a regulated environment? I'd like to hear how you are approaching it. Find me on LinkedIn or take the free frameworks.